Eyesis4Pi Calibration

From ElphelWiki
Revision as of 10:27, 22 December 2016 by Oleg (talk | contribs)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search

Goniometer and IMU

Fig.1 Positive rotation angles in the calibration machine
Fig.2 Goniometer
Fig.3 IMU in Eyesis4π

IMU is ADIS 16375 - 2400 samples per second

During the calibration process the goniometer rotates the camera about 2 axes (y & z on Fig.1) by precisely known angles. To calibrate IMU one needs to match the goniometer angles with the IMU data. Notes:

  • The images are taken in the middle of the rest state of the goniometer.
  • IMU log contains 3 types of events: IMU data packet, GPS data packet, Image captured.
  • Images timestamps are recorded in the IMU log.

Calibration process output examples:

Camera 00:0E:64:08:1B:89 IMU calibration data:

Axial - angle about the z-axis in degrees (positive rotation is shown on Fig.1)
Tilt - angle about the y-axis in degrees (positive rotation is shown on Fig.1)